Abstract

The pressure and temperature dependent fluorescence line-shift of strontium tetraborate has been measured concurrently with x-ray diffraction from the pressure standards sodium chloride or gold. Temperature was found to have a small effect on the fluorescence line-shift under pressure. We found a maximum pressure uncertainty of ±1.8 GPa at 25 GPa (7.2%) and 857 K when making no temperature correction. The fluorescence line-shifts for ruby, Alexandrite, and samarium-doped yttrium aluminum garnet were also determined, using our strontium tetraborate calibration to determine pressure and a thermocouple to measure temperature. Fluorescence measurements were extended up to 800 K for ruby and Alexandrite. Temperature was found to have a small effect on the fluorescence line-shift of samarium-doped yittrium aluminum garnet. We found a maximum uncertainty of ±2.7 GPa at 25 GPa (11.1%) and 857 K when no temperature correction was applied. We determined equations relating to the fluorescence line position from these data, which include a cross derivative term to account for the combined effect of pressure and temperature. We present a method to independently determine pressure and/or temperature from combined fluorescence line-shift measurements of a pair of optical sensors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call