Abstract

Measurement of theR wave amplitude and evaluation of the sensitivity reserve of implantable apparatus during cardiomyoplasty make it possible to prevent postoperation desynchronization of the electrostimulator. The sensitivity reserve for an ECS-445 electrostimulator is established. Measurements of acuteR waves in 15 patients are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.