Abstract

With the goal of enhancing the low sensitivity of the conventional atmospheric pressure chemical ionization (APCI) source due to the ion divergence, here is reported a T-shaped ion source to improve the sensitivity with a portable mass spectrometer for the rapid determination of the pesticides ethoprophos and atrazine in water. Compared with conventional APCI, the ion source improves the sensitivity for the analytes. The results show that the analysis requires less than 2 s, the limit of detection (LOD) is better than 0.1 and the relative standard deviation (RSD) is 16% at 1, 5, and 25 The recoveries of the analytes are between 80% and 130%, meeting the requirements for rapid pesticide determination in water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call