Abstract
In this work the detection limits of arsenic, zinc and lead in silica matrix by X-ray photoelectron spectroscopy (XPS) analysis were evaluated. These elements were selected since they are particularly relevant in environmental investigations. Pure oxides and several binary mixtures with silica at different concentrations of As (III), Zn (II) and Pb (II) oxides obtained by ball milling were analysed by XPS. Inductive-coupled plasma optical emission spectroscopy (ICP–OES) was also exploited for determining their bulk concentration. The linear response of XPS intensity vs weighed (at%) was confirmed. Three different models were here applied to estimate the limit of detection (LOD) of arsenic, zinc and lead and the results were in good agreement. Under the experimental conditions here adopted, the calculated LODs (at%) of arsenic, zinc, and lead as pure oxides and in silica matrix were found to range between 0.1 and 0.3, between 0.04 and 0.1 and between 0.03 and 0.04 respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.