Abstract Specific directions for the thermistor cryoscopic method, including uniformity in cooling, degree of supercooling, seeding, and reading procedures, are proposed for the present official method. It is emphasized that the instrument must be in the proper condition for use, must be properly calibrated, and must be properly utilized by the analyst. Each analyst should individually calibrate with standards by the same uniform procedure for standards and milk samples. The precautions necessary in the determination are part of the method and must be observed if the same sample is to yield the same freezing point value for different analysts, in different laboratories, at different times