Abstract
Ex situ plated Bi film electrodes (Bi-FE) have been employed, for the first time, to measure the free concentration of Pb(II) in aqueous solutions using absence of gradients and Nernstian equilibrium stripping (AGNES) with stripping chronopotentiometry (SCP) quantification. The amount of deposited Pb°, below a certain threshold, follows a Nernstian relationship with the applied potential. This threshold can be interpreted as the frontier of transition from surface deposition to solid (bulk) formation of Pb°. AGNES with Bi-FE yielded a very good detection limit (3σ) for Pb(II) of 6.0 × 10(-9) M with an applied gain of 398 and a deposition time of 400 s. The ability of the Bi film electrode to perform speciation measurements was demonstrated for Pb(II)-PSS and Pb(II)-IDA systems. The measured values with the Bi-FE were in good agreement with the values obtained using the Hg film electrode and/or the values reported in the literature.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.