Abstract
Objective To establish a method for the determination of acyclovir in human plasma. Moth380nm. The column temperature was 30℃ and the sample size was 20 μl.Results The linear range of acyclovir was 112~1120ng/ml (r=0.9999),the methods recovery was 96.85%~99.24%,withinday-day and between-day RSD was less than 3%.Conclusions The method is simple,accurate,special and can be used for the pharmacokinetics study of acyclovir. Key words: RP-HPLC; Acyclovir; Plasma concentration; Fluorescence detector
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.