Abstract

CR-39 based radon detectors are widely used in measuring indoor radon. In this regard, different groups have developed their own systems. However, before using any system for indoor radon measurements, it has, first, to be calibrated with a known source of radon. In the current study, CR-39 based NRPB type radon detector has been calibrated and presented. In this regard, about 200 holders for CR-39 were obtained from the Radiation Protection Division of the Health Protection Agency (former NRPB), UK and several thousand more similar detector holders, hereafter called NRPB type holders, were fabricated locally in Pakistan. Uranium ore samples of known grade were placed into the plastic containers of volume 5.4 × 103 cm3 and CR-39 detectors were placed in the NRPB type holders and were then installed into the containers at a distance of 25 cm from the surface of the known grade ore samples. The containers were hermetically sealed and the detectors were allowed to expose to radon for 3 weeks. After 16 h etching in 25 % NaOH at 80 °C, the measured track densities were related to the radon concentration. The calibration factor of 2.563 tracks cm−2 h−1/kBq m−3 was obtained.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.