Abstract

The mcroLion chamber, which has a small volume (0.0017cc) and a high sensitivity suitable for small field dosimetry, is an ionization chamber with liquid (isooctane) filling the cavity volume. In the present work, we used two alternative ways to estimate the beam quality correction factor \(k_{Q,Q_0 }^{micL} \) of the microLion chamber in the 6 MV photon beam from a Clinac iX accelerator (Varian, USA); Monte Carlo simulations and experimental methods. The Monte Carlo calculation (1.0183 ± 0.58%) agreed to within 0.56% with the experimental determination (1.024 ± 0.58%). The results from the two methods were in good agreement within acceptable uncertainties. We think that the averaged factor, 1.0212, could be used for the microLion chamber in reference dosimetry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.