Abstract

A differential-pulse adsorptive stripping voltammetric method for the determination of trace amounts of the antibacterial trovafloxacin (TRFLX) is proposed. The optimal experimental parameters for the drug assay were: accumulation potential=−0.30 V (vs. Ag/AgCl), accumulation time=120 s, pulse amplitude=50 mV and scan rate=5 mV s −1 in Britton–Robinson buffer (pH 4.5). The linear concentration range of application was 2.0–20.0 ng ml −1 of TRFLX, with a relative standard deviation of 3.6% (for a level of 5.0 ng ml −1) and a detection limit of 0.6 ng ml −1. The method was applied to determination of TRFLX in human urine and serum samples. It was validated using HPLC as a reference method. Recovery levels of the method reached 100% in all cases

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.