Abstract

An enantioselective method for the separation and quantification of the diastereomer pairs of metolachlor and S-metolachlor in surface and ground waters is presented. Samples are purified and concentrated using a C18 (octadecyl silica) solid-phase extraction (SPE) procedure and analyzed by chiral column liquid chromatography-mass spectrometry/mass spectrometry (LC/MS/MS) interfaced with either atmospheric pressure chemical ionization (APcI) or atmospheric pressure photoionization (APPI) sources. The overall mean percent procedural recoveries (percent relative standard deviations) are 89% (10.6%) for surface water and 80% (9.1%) for ground water. The method limit of quantitation (LOQ) is 0.10 ppb. The method validation was conducted under U.S. EPA FIFRA Good Laboratory Practice Guidelines 40 CFR 160.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.