Abstract
Objective: To instruct a method of determining thallium in the urine by graphite furnace atomic absorption spectrometry(GF-AAS) with colloidal palladium as the matrix modifier. Methods: Urine samples were first diluted and then determined by GF-AAS with colloidal palladium while using thermal sample injection. Results: The optimum volume of colloidal palladium was 6 μl and the best ashing temperature was 600-800 ℃ while the atomization temperature was 1700-1900 ℃ . This method showed a good linearity relationship when the concentration between 0.33 and 50.0 μg/L while the correlation coefficient of standard curve line was 0.9992, and the detection limit was 0.33 μg/L and the recovery rate was between 92.7% and 102.3% with the intra-day precision in the range of 2.55% to 3.66% and the inter-day precision in the range of 1.77% to 3.85%. Conclusion: This method has the advantages of low detect limit, high sensitivity and good precision, and it can be used in the biological monitoring and emergency detecting of workers exposed to thallium.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Zhonghua lao dong wei sheng zhi ye bing za zhi = Zhonghua laodong weisheng zhiyebing zazhi = Chinese journal of industrial hygiene and occupational diseases
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.