Abstract

A novel analytical technique for the determination of tetramethylthiuram disulfide (TMTD) was developed. It is based on total oxidation of TMTD with hypobromite. Hypobromite is generated on one segment of interdigitated array. It reacts rapidly and quantitatively with TMTD and unreacted flux of hypobromite is detected on second segment of the interdigitated microelectrode array (IDA). Detection is amperometric therefore the potential of the second (detector) segment is fixed at the value corresponding to the limiting current of the opposite reaction to that taking place on the generator segment. Rather low detection limit of this method ( 7×10 −7 mol dm −3 ) allows the analysis of samples with trace content of TMTD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.