Abstract

To establish an ultra-high performance liquid chromatography coupled with triple quadrupole mass (UPLC-TQ-MS) for determination of four terpene lactones. Chromatographic separation was carried out on a ACQUITY UPLC BEH C18 column (2.1 mm x 100 mm, 1.7 microm) with isocratic elution of 70% methanol at a flow rate of 0.4 mL x min(-1), the column temperature was set at 30 degrees C; Waters Xevo TQ worked in multiple reaction monitoring mode. All calibration curves were linear (r > 0.990 3) over the tested ranges. The average recoveries ranged from 98.83% to 103.9% with RSD value below 3.0%. The contents of total terpene lactones in Ginkgo biloba leaves were significantly different in different ages. The contents in the leaves of young ginkgo tree were higher than that in old tree. The method was simple and fast with high precision, sensitivity and repeatability, which can be used for qualitative and quantitative analysis of terpene lactones in G. biloba leaves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.