Abstract

A sensitive high-performance liquid chromatography method for the determination of taurine in human plasma was developed. Taurine and N-methyltaurine (internal standard) were derivatized with 4-(5,6-dimethoxy-2-phthalimidinyl)-2-methoxyphenylsulfonyl chloride to produce fluorescent sulfonamides. The labeling reaction was carried out at 70 °C for 20 min at pH 7.5. The fluorescent derivatives were separated on a reversed-phase column by a stepwise elution using (A) acidic phosphate buffer/acetonitrile (83/17) and (B) acetonitrile and detected by fluorescence measurement at excitation and emission wavelengths of 318 and 392 nm, respectively. The detection limit (signal-to-noise ratio=3) of taurine was 3 fmol per injection. The within-day and day-to-day relative standard deviations were 3.0–4.8 and 2.5–4.7%, respectively. The concentration (means) of taurine in normal human plasma was 48.9 ± 7.5 μM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.