Abstract

To establish a derivative fluorometry method for the determination of sulfur dioxide residues in traditional Chinese medicine. The optimal derivation condition was established. The fluorescence intensity was detected at excitation wavelength of 321 nm, and emission wavelength of 384 nm. A linear relationship was obtained between the fluorescence intensity and the addition of reference substance in the range of 0.999 7-17.99 nmol with a correlation coeffient of 0.999 9, and the average recovery was 102.3% with RSD 4.6%. This method is simple and sensitive with quick and correct result. It can provide a reference for the determination of sulfur dioxide residues in traditional Chinese medicine.