Abstract
An automatic thermodenuder system for the measurement of SO/sub 2/ is described. Air is sampled by Cu/CuO coated denuder tubes. When the denuders are heated to 800/sup 0/C, SO/sub 2/ is liberated and detected by a standard monitor. For a sampling time of 30 min the limit of detection is 0.1 and 0.015 ..mu..g/m/sup 3/ SO/sub 2/ for a pulsed-fluorescence and a flame-emission monitor, respectively. For both instruments the response is linear up to 100 ..mu..g/m/sup 3/. Sulfuric acid, sulfate aerosols, and organic sulfur compounds like CS/sub 2/, COS, and dimethyl sulfide do not interfere at concentrations up to 100 ..mu..g/m/sup 3/. Interference of H/sub 2/S can be eliminated by a NaHSO/sub 4//Ag/sub 2/SO/sub 4/ coated predenuder placed in the sample line. The measurement of SO/sub 2/ concentrations at background levels and the determination of dry deposition velocities by measurement of SO/sub 2/ gradients are obvious applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.