Abstract
The capabilities of the ligand-exchange capillary electrophoresis mode (working electrolyte, 1 mM copper(II) and 175 mM NH3; 30 mbar, 10 s; +25 kV; 245 nm; 20°C) for the determination of sugars (glucose, fructose, and sucrose) in real samples (fruit juices, wines, and drug preparations) were studied. With the use of glucose as an example, the detection limits for direct (as sugar—copper(II) complexes; 245 nm) and indirect (working electrolyte, 5 mM tryptophan and 50 mM NaOH; 30 mbar, 10 s; +10 kV; 280 nm; 20°C) UV detection were compared: ≈21 and ≈8 mg/L (for direct and indirect detection, respectively).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.