Uranium(VI) is selectively determined by a compleximetric titration with pyridine-2,6-dicarboxylic acid, using arsenazo-I indicator and hexamethylenetetramine buffer at pH 4.9. Cyclohexanediaminetetraacetic acid and diethylenetriaminepentaacetic acid provide masking of interfering metal ions. A probe colorimeter apparatus is recommended for end-point detection. The relative standard deviation is 0.6% for 0.17–0.76 μmol of uranium.