Abstract

This case study describes the experimental determination of displacements and stresses on a composite model of a pelvis that was modified to represent a healthy intact pelvic ring. The modified model was stressed statically up to 1750 N to simulate standing on one leg and also cyclically to model walking. For two different model settings in the loading machine the values of displacements and stresses at the pelvic ring were determined. The two different settings correspond to two different loading vectors applied on the pelvic ring, boundary conditions and degrees of freedom. The experimentally determined values of displacements in both settings are very similar and in accordance with the knowledge on the behaviour of a real human pelvis. The modified model is thus suitable for testing of newly developed implants for pelvis treatment and experimental determination of displacements and stresses in pelvic ring which are caused by application of implants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.