Abstract

This paper discusses the technique for high-precision quantification using (1)H-NMR to determine the purity of analytical standard samples. The procedure described is based on the use of internal reference samples in an (1)H NMR experiment in our laboratories. The sample preparation and all relevant NMR parameters were optimized for minimum uncertainty. The validation of accuracy and precision was performed by comparing different certified reference materials. It was shown that the high-precision measurement is applicable even for relatively small sample amounts down to 2.5 mg. The relative combined uncertainty of measurement was found to be 0.15%. Two different approaches for uncertainty calculation were compared; a complete uncertainty budget was calculated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.