Abstract
The specific heat capacities (cp) for the brain and muscle equivalent liquids were determined with isothermal heat conduction microcalorimetry (IMC) and differential scanning calorimetry (DSC). IMC was found to afford an accurate technique to measure cp for solid and liquid samples, when an appropriate reference is employed. The accuracy of obtained cp values was estimated to be better than 0.7% with the equivalent liquids. Intercomparison with a conventional isoperibolic calorimeter showed an excellent agreement within the estimated uncertainty of the isoperibolic calorimeter (±3%). Additionally, suitability of different kinds of IMC sample vessels was tested, and the standard electrical calibration procedure of IMC was evaluated through the determination of cp with and without a reference material.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.