Abstract

The objective of this paper is to report a procedure for the extraction and the determination of five textile dyes (Lanasyn Blue F-L 150, Lanasyn Dark Brown M-GLN, Lanasyn Red M-GA, Nylosan Dark Brown S-MBL, and Nylosan Red N-2RBL) which were not yet studied in literature by high-performance thin-layer chromatography. For SPE of these dyes, four types of Strata cartridges (WAX/NH2, SAX, C18-U and C18-E) were tested. The best results were obtained on Strata WAX/NH2. The dyes were separated on precoated Alugram RP-18W/UV254 plates with the n-butanol-ethyl acetate-5% ammonium hydroxide 4:4:1 (v/v) mobile phase. The densitometric scanning was performed at 550 nm. Linear regression analysis of the calibration data indicated a good linear relationship between the peak area and the dye concentration in the range of 20–60 ng per band. For the HPTLC method, the detection and the quantification limits of the studied dyes were calculated. The obtained detection limits (ng per band) were: 6.6 for Lanasyn Blue F-L 150...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.