Abstract
The study aimed to determine appropriate detectors for output factor measurement of small fields in 6 and 10 MV flattening filter free photon beams using five different detectors. Field sizes were varied between 0.6 × 0.6 and 4.0 × 4.0 cm2. An indirect method (daisy-chaining) was applied to normalize the output factors. For the smallest field size, the variations of output factors compared among the detectors were 13%. Exradin A16 had the lowest output factor and increasing in sequence with CC01, microDiamond, microLion and EDGE detectors, respectively, for both energies. The similarity between CC01 and microDiamond output factor values were within 1.6% and 1% for all field sizes of 6 and 10 MV FFF, respectively. EDGE and microLion presented the highest values while ExradinA16 gave lowest values. In conclusion, IBACC01, Exradin A16, microLion, microDiamond and EDGE detectors seem to be the detectors of choices for small field output factor measurement of FFF beams down to 1.6 × 1.6 cm2. However, we could not guarantee which detector is the most suitable for output factor measurement in small field less than 1.6 × 1.6 cm2 of FFF beams. Further studies are required to provide reference information for validation purposes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.