Abstract
A photon correlation spectroscopy method has been developed to characterize the size distribution of fat globules in intravenous fat emulsions (IFE) in terms of mean diameter, standard deviation of the distribution, and percentage of large particles outside the distribution. Mean fat globule diameters of samples of all IFE products available in Canada were about 0.3 microns, similar to values reported in the literature. The methodology is sufficiently sensitive to detect the presence of 5% by weight of 2 microns polystyrene microspheres in an intravenous fat emulsion. The effect of changes in instrument settings and variables on the results has been evaluated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.