Abstract

A method for the determination of silicon in titanium dioxide and zirconium dioxide powders based on electrothermal atomic absorption spectrometry using the slurry sampling technique has been developed. The experimental conditions with regard to chemical modification, the temperature programme, the lifetime of the graphite furnace and the slurry concentration were optimized. The behaviour of zirconium and calcium in the graphite tube during the ashing, atomization and cleaning steps was investigated using 47Ca and 97Zr as radiotracers. Calibration was performed by the standard additions method using aqueous standards. The results of this technique were compared with those for atomic emission spectrometry. The limits of detection were found to be 7 µg g–1 in titanium dioxide and 2 µg g–1 in zirconium dioxide, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.