This study aimed at developing and validating an HPLC method for the assay of sildenafil citrate and its related substances that might coexist in the drug commercial products and in tablets' formulation as impurities that originate from synthesis processes or degradation. A chromatographic system comprising a microBondapak C(18) (10 microm) column, a mobile phase of ammonium acetate (pH 7.0, 0.2 M)-acetonitrile (1:1, v/v), a flow rate of 1 ml/min and a UV detector set at 240 nm has shown good chromatographic separation for sildenfil and the other related substances. The degree of linearity of the calibration curves, the percent recoveries of sildenafil and related substances, the limit of detection, LOD, and limit of quantitation, LOQ for the HPLC method have been determined. The HPLC method under study was found to be specific, precise, accurate, reproducible indicating stability and robust.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.