Abstract

Background: In clinical laboratories, Determination of serum testosterone levels in routine is challenging. Reliable analysis of bioavailable testosterone is difficult. Methods: After liquid-liquid extraction, Shimadzu Prominence LC unit coupled to API 3200 mass spectrometer with atmospheric pressure chemical ionisation was used to quantify serum testosterone levels. Serum testosterone results taken from tandem mass spectrometry were compared with Roche, Beckman and Abbott commercial automated testosterone immunoassays Results: In LC-MS/MS method, linearity for testosterone was 0.005-15.915 ng/mL. Recovery was 97-108% for 0.06-16 ng/mL concentration range. Intraday and inter-day CVs were 6.3,9,7; 2,2,8.7 and 1.3,0.3 for low (0.969 ng/ mL), intermediate (4.816 ng/mL) and high (8.084 ng/mL) testosterone concentrations, respectively. There was a good correlation between commercial immunoassays. In method comparision, agreement was closest with Roche and LCMS/MS method (r=0.9670). Discussion: In contrast to commercial testosterone immunoassays, LC-MS/MS method is a reliable, simple and useful method for clinical laboratory practices.