Abstract

A liquid chromatographic-mass spectrometric method for the determination of cortisol in serum using atmospheric pressure electrospray ionization and tandem mass spectrometry is described. During sample preparation, 150 microl of serum were deproteinized with methanol/zinc sulfate followed by on-line solid phase extraction employing column switching. Tri-deuterated cortisol was used as the internal standard. The following transitions were monitored: cortisol, 363>309 m/z; d3-cortisol, 366>312 m/z. The total run-time was 5 minutes. The method proved linear (0-500 microg/l; r=0.999), precise (total coefficient of variation between 5.0% and 3.2% at a mean cortisol concentration of 15.1 microg/l and 269 microg/l, respectively; n=16) and specific with regard to relevant endogenous and exogenous steroids.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.