ABSTRACTObjectives: Carbamazepine (CBZ) is a commonly used anti-epileptic in rural hospitals in India. These hospitals lack the facilities to measure CBZ concentration; however, in larger hospitals this is performed using high performance liquid chromatography (HPLC). Dried blood spot (DBS) represents a feasible matrix for safe transportation by post/courier. This study was to determine whether the concentration of CBZ in serum can be predicted from that measured in DBS using an inexpensive HPLC method and inexpensive standard filter paper.Methods: CBZ in serum and DBS from 80 epileptic patients were measured using a validated HPLC assay. The data was then randomly divided into two groups; simple Deming regression was performed with the first group and validation was performed using the second.Results: There was a good correlation between the serum and DBS concentrations (r = 0.932) in the first group. The regression equation obtained was: predicted serum concentration = DBS concentration x 0.83 + 1.09. In the validation group, the correlation between the predicted and actual serum concentrations was also good (r = 0.958), and the mean difference between them was only 0.28 μg/ml (p = 0.8062). The imprecision and bias in both the groups were acceptable.Conclusion: Using inexpensive materials, serum CBZ concentrations can be accurately predicted from DBS specimens. This method can be recommended for the therapeutic drug monitoring of CBZ in resource-limited settings.