Abstract

A 6% slurry of polytetrafluoroethylene (PTFE) in 4% hydrofluoric acid and 1% nickel nitrate were used as modifiers for determination of selenium in soils by GF AAS. PTFE was used to remove silica from the soil sample, because this resulted in severe matrix effects. The temperature of fluorination, determined thermogravimetrically, was 600 degrees C. The yield of fluorination depends on the molar ratio of PTFE/silica, particle size, and the time and temperature of fluorination. The soil samples were pretreated with a small amount of concentrated hydrofluoric acid placed directly in the cup of autosampler. The results for the determination of selenium in the reference soil materials by means of the slurry-sampling technique and use of aqueous standards are in good agreement with the certified values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.