Abstract
Abstract The essential trace element selenium is a focus of attention due to its effects on human health, there being consequences of both its deficiency and excess. Due to the ultra-trace content of selenium, the neutron activation analysis method (NAA) is difficult to apply. We therefore made use of the radiochemical neutron activation analysis (RNAA) to determine Se at low level concentrations in several consumed food items in Algeria. A radiochemical procedure based on liquid–liquid separation was established in our laboratory. In this research we focused on the determination of selenium in two species of coffee: Arabica and Robusta. The accuracy of the method was assessed by analyzing the certified reference material NIST-SRM 1573a (tomato leaves). The results obtained show a selenium variation from 0.025 to 0.052 μg/g in coffee beans and an average yield of the separation of about 85%. The results of this study were compared with those obtained with samples from Brazilian, Caribbean, Indian and Kenyan coffee beans.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.