Abstract

On-line removal of transition metal interferences in microscale suction-flow hydride generation atomic absorption spectrometry is described for the determination of selenium at μg l −1 levels. A mini-column of a chelating resin with iminodiacetate groups is used. Selenium in solutions containing ⩽ 2.5 mg of copper or nickel was determined at a rate of 30 samples per hour; the detection limit was 0.1 ng of selenium. The relative standard deviation for ten replicate measurements of 5 ng of selenium was 3.8%. The method was applied successfully to the determination of selenium in standard copper alloys and nickel sponge.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.