Abstract

A quantitative assay for scutellarin by LC–MS–MS (negative ion mode) was developed. The scutellarin was extracted from dry Erigeron breviscapus. Significant ion suppression was observed, which could be eliminated by increasing the turboionspray interface temperature to 350°C and by 1000-fold dilution of the extract with solvent. The calibration curve of scutellarin showed excellent linearity over a wide concentration range (0.01–100 μg/ml) ( r=0.998), and the limit of detection was 15 pg/ml using a 10-μl injection volume. The analysis time was 4 min/sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.