Abstract
AbstractA semi‐empirical approach is described for determining the mass per unit area of a sample being analysed. The method can be used to estimate the concentration of minor and trace elements in matrices containing a substantial amount of light elements. The procedure utilizes the coherently and incoherently scattered radiation induced in the sample by the filtered continuum radiation of a rhodium x‐ray tube. The relationship between the intensity of the scattered radiation per unit mass and the average atomic number of the sample is established via calibration graphs, which can be applied for different x‐ray tube voltages and for different primary beam filters. The overall procedure was validated by the analysis of several geological standards, deposited as thin slurries of unknown thickness either on Mylar foil or on Nuclepore filters.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.