Abstract

A selective, sensitive and low sample consumption GC method for determination of residual solvents in PSMA samples was developed. Selectivity with respect to solvents such as ethanol, diethyl ether, 2-propanole, acetonitrile, dichloromethane, used in manufacturing of PSMA-11, was obtained on two different GC systems and columns. In all these cases, good separation and other system suitability parameters were obtained. The values of detection limit (LOD) and limit of quantitation (LOQ) were estimated and verified experimentally for ethanol, diethyl ether, 2-propanol (IPA), acetonitrile (ACN) and dichloromethane. Full validation parameters i.e.: selectivity, limit of detection (LOD), limit of quantitation (LOQ), linearity, precision, accuracy, robustness and system suitability parameters are presented with respect to ACN and IPA. Accuracy and precision were measured on two GC systems. Statistical tests exhibit, that there are no significant differences in accuracy and precision between two GC systems

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.