Abstract

The cationic polyelectrolyte pDADMAC is widely used in biopharmaceutical industry as a flocculating agent to enhance clarification throughput and downstream filtration operations. Due to the possible toxicity, pDADMAC should be assessed for an acceptable residual level to ascertain the safety of the product to patients. The strong protein-polyelectrolyte interaction, however, can negatively affect sensitivity and accuracy of measurements. This paper reports on the application of size exclusion (SE) chromatography coupled to evaporative light scattering detector (ELSD) to the quantitative determination of pDADMAC in monoclonal antibody formulations and in process intermediates during downstream purification. The SE chromatography was performed under isocratic condition with a mobile phase consisting of 0.1% TFA in water (90%) and acetonitrile (10%) at a flow rate of 0.4 ml/min. A quantification limit (S/N = 10) of 0.85 ppm was achieved in sample matrix, which is sufficiently low for the trace analysis of this compound in protein-containing samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.