Abstract

A method for the direct determination of paraquat, diquat, chlormequat and difenzoquat in water samples, using an on-line solid-phase extraction–liquid chromatography–tandem mass spectrometry system was developed. No sample preparation was required and the detection limits were below the European Union maximum residue levels. The chromatographic separation was performed using an XTera MS C 8 column. The concentration of the ion pair reagent, the pH and the gradient elution were optimized to give high recoveries and good chromatographic resolution between quats. The detection was carried out using an ion trap as mass analyzer. Parameters such as the magnitude and duration of the resonant excitation voltage and the magnitude of the trapping RF voltage for full scan tandem mass spectrometry (MS–MS) experiments were studied to establish the optimal experimental conditions. Moreover, the accurate optimization of these parameters allowed MS–MS experiments of low mass ions, below m/z 200, providing unambiguous peak identification. Finally, the reproducibility of the proposed method was shown by good run-to-run and day-to-day precision values and its applicability to the determination of quats in drinking water was evaluated using spiked samples.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call