Abstract

Platinum in coke samples was determined by inductively coupled plasma mass spectrometry (ICP-MS) after microwave-assisted sample digestion and temperature controlled microwave assisted cloud point extraction (MW-CPE). Interferences due to hafnium (179Hf16O) were successfully eliminated using MW-CPE with 2-MBT prior to the determination of platinum by ICP-MS. The validity of the method was studied by spike recovery tests and by analyzing certified reference material (BCR-723 street dust). The results obtained for platinum in the BCR-723 were generally in good agreement with the certified values. Furthermore, the preliminary platinum results obtained for coke sample by ICP-MS after MW-CPE were compared to those obtained by a reference method (NiS-Fire Assay preconcentration/Te coprecipitation and ICP-MS determination).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call