Abstract

This paper describes a simple and efficient procedure for measuring 25 congeners of polychlorinated biphenyls in human milk. The limit of quantitation was 0.1 ng/ml for five less chlorinated congeners (PCB 70, 74, 87, 99,101), and 0.01 ng/ml for the remaining 20 congeners (PCB 77, 105, 118, 126, 128, 138, 151, 153, 156, 169, 170, 180, 183, 187, 191, 194, 205, 206, 208 and 209). Solid phase extraction technology was applied to extract the analytes from the matrix and to remove lipids. Three columns were used sequentially, and they were a Bond Elut C 18, a Sep-Pak Plus NH2 and a Bond Elut PCB cartridge. The instrumental method was gas chromatography–mass spectrometry with negative chemical ionization, and selected ion monitoring mode was used.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.