Abstract

A reliable and sensitive method for the determination of plutonium in environmental samples by a combined use of alpha-spectrometry and accelerator mass spectrometry (AMS) is presented. Different environmental samples contaminated with plutonium from the Chernobyl accident and other sources were investigated using both methods. Plutonium was chemically separated using extraction chromatography. As a chemical yield tracer, 242Pu was applied for both AMS and alpha-spectrometry. After the alpha-spectrometric measurements, the samples were converted to targets suitable for the AMS measurements. AMS was especially chosen in addition to alpha-spectrometry, because it allows the determination of 240Pu/239Pu isotopic ratios as well as detecting the beta-emitting radionuclide 241Pu. Besides the generally used activity ratios of 238Pu/239,240Pu, there are now further optional isotopic ratios available using this method. By combining AMS with alpha-spectrometry, all relevant plutonium isotopes can be measured. Thereby a reliable distinction of different sources of plutonium in the environment is possible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.