Abstract

A measurement procedure for determining of Ru, Pd, Ir, Pt and Au mass fractions in ferromanganese deposits by inductively coupled plasma‐mass spectrometry after acid digestion and anion exchange preconcentration is presented. To eliminate incomplete recovery after sorption preconcentration of the platinum‐group elements (PGE) and Au, a standard addition method was used. Detection limits ranged from 0.02 ng (Pd, Ir) to 0.19 ng (Ru). The measurement results for ferromanganese nodule reference material NOD‐A‐1 and NOD‐P‐1 agree with earlier reported values. Intermediate precision of PGE concentration data for nodule reference materials in this work was 5–24% (1s) and could reflect sample heterogeneity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.