An inductively coupled plasma atomic emission spectrometry (ICP-AES) method was developed for the determination of phosphorus in fertilizers. Total phosphorus, direct extraction available phosphorus (EDTA), and water-soluble phosphorus, reported as phosphorus pentoxide (P205), in 15 Magruder check fertilizers were measured by ICP-AES, and the results were compared with those obtained by the AOAC official method. Five analytical wavelengths of phosphorus, 177.499, 178.287, 213.618, 214.914, and 253.565 nm, were tested for the determination of phosphorus in fertilizers, and their detection limits were obtained. Acid effects of perchloric acid and possible matrix effects of aluminum, calcium, magnesium, potassium, and sodium were negligible for phosphorus determination. Wavelength 213.618 nm was the best analytical wavelength for phosphorus determination by all 3 sample preparation methods for the selected Magruder fertilizers. The results demonstrated that the accuracy and precision of the ICP-AES method were comparable with those of the official methods.