Abstract

A modified molybdenum blue method for fast and accurate measurement of arsenate and phosphate in aqueous solution at concentrations below 10 μmol l −1 is reported. The modification consists of optimizing the composition of the molybdenum-containing solution (potassium antimony tartrate, ammonium molybdate, sulfuric acid). Selective reduction of arsenate by sodium dithionite is used to determine phosphate concentrations, and for the speciation of arsenate and arsenite, in an aqueous mixture according to the scheme developed by Johnston and Pilson. Sodium dithionite is used for the first time to achieve complete, fast (<10 min), and selective reduction of arsenate in neutral solution. These two significant improvements afforded a colorimetric limit for As detection near 1 ppb, which easily meets the requirements imposed by the revised EPA threshold levels for As in drinking water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.