Abstract

Headspace solid phase micro-extraction (HS-SPME) has been applied as a sampling technique for the determination of petroleum contamination in shellfish using capillary gas chromatography-mass spectrometry (GC-MS). A poly(dimethylsiloxane) fused silica fibre (100 microm thickness) was found to be satisfactory for the extraction of a range of aliphatic hydrocarbons (HCs) from homogenised shellfish tissues. The SPME conditions, including temperature, salt content, extraction time and desorption temperature, were optimised for a range of aliphatic HCs (C9-C20). A methyl silicone column GC (12 m x 0.20 mm, 0.33 microm layer thickness) was used with a temperature programme from 40 to 260 degrees C and the HCs were determined within a mass range of m/ z=50-550 in electron impact mode. Calibration range was from 10 to 5000 ng/g with linear correlation coefficients ( r(2)) of 0.982 for nonane to 0.997 for octadecane. Detection limits for aliphatic HCs, spiked into shellfish (mussel) tissues, varied from 3.6 ng/g (tetradecane) to 51 ng/g (eicosane) and relative standard deviation (% RSD) values ranged from 1.4% (hexadecane) to 24.3%(eicosane).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.