Abstract

A systematic study comparing the methodology and analytical results obtained in an investigation of seven pesticide residues (Molinate, Atrazine, Carbofuran, Pirimicarb, Prometryn, Malathion and Tetrachlorvinphos) in soil samples is reported. Solid-phase extraction (SPE) using glass columns and 47 mm disks of octyl and octadecyl-bonded silica was used in the pesticide analysis. The best extraction efficiency and clearest extracts are obtained with C8 disks. The analyses were carried out by capillary gas chromatography with nitrogen and phosphorus detection. Recovery experiments were performed at ppb levels in spiked soil samples. The average recoveries of the compounds were 53–77%. Detection limits are between 5 and 30 ng g−1 based on 5 g moist soil sample. The method was validated by comparing it with conventional liquid-liquid extraction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.