Abstract

A comparative study on the use of cetyltrimethylammonium bromide (CTAB)-coated silica and sodium dodecyl sulphate (SDS)-coated alumina mixed hemimicelles-based solid-phase extraction (SPE) for the pre-concentration of six perfluorinated compounds (PFCs) in environmental water samples was presented. The six analytes heptafluorobutyric acid (HFBA), perfluoroheptanic acid (PFHeA), perfluorooctanic acid (PFOA), perfluorooctanic sulfonic (PFOS), perfluorononanic acid (PFNA) and perfluorodecanic acid (PFDeA) were quantitatively retained on both sorbent materials. The cationic surfactant (CTAB adsorbed onto silica) was more appropriate for SPE of PFCs. The main factors affecting adsolubilization of PFCs including the amount of surfactant, pH of solution, sample loading volume and desorption were investigated and optimized. Concentration factor of 500 were achieved by SPE of 500 mL of several environmental water samples. The method detection limits obtained for HFBA, PFHeA, PFOA, PFOS, PFNA and PFDeA were 0.10, 0.28, 0.07, 0.20, 0.10 and 0.05 ng/L, respectively. The relative standard deviation of recoveries ranged from 2 to 8%, which indicated good method precision.