Abstract

Perchlorate accumulation in flame atomic absorption systems has been studied using ion-pairing chromatography with indirect photometric detection. The results indicate that highly variable amounts of chloride are produced during the aspiration of perchlorate salts. Perchlorate accumulation was found to be related to the species aspirated and to the burner and spray chamber design. The effects of flame stoicheiometry, rinsing and preventative additives are discussed. The chromatographic technique and experimental design may assist in improving atomic absorption systems in order to provide optimised safety with respect to perchlorate-induced spray-chamber explosions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.