Abstract
A particle‐induced X‐ray emission (PIXE) technique has been used in the determination of the principal components Pb, Zr and Ti and the substituting elements Sr, Cr, Nb and La in lead zirconate titanate ceramics. In general, precision of analysis was concentration dependent from each element under study. For Pb, precision varied between 0.13% and 0.16%, at higher concentration of 59.32–64.5%. It was around 6–9% for Sr, Cr and Nb at concentrations of 1% or lower. Particular attention was devoted to the estimation of the analysis trueness. With this purpose, three methods were applied: (1) comparison of PIXE and inductively coupled plasma optical emission spectrometry concentrations, (2) recovery study and (3) comparison with a laboratory standard. Trueness of analysis was around 100 ± 10% for the evaluated elements Pb, Zr, Ti, Sr and Cr. The expected stoichiometry and elemental composition homogeneity of a wide group of produced ceramics were confirmed by the PIXE technique. Copyright © 2012 John Wiley & Sons, Ltd.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.