A simple and efficient procedure for separation and pre-concentration using ultrasound-assisted co-precipitation with manganese dioxide was developed for Pb determination by inductively coupled plasma optical emission spectrometry (ICP OES). The optimization process was carried out using a two-level factorial design and a Doehlert matrix. Three variables (i.e. concentration of oxidizing solution—KMnO 4, concentration of MnSO 4 solution and time of ultrasonic irradiation) were used as factors in the optimization. The recoveries, based on the analysis of spiked samples, were between 90% and 105%, and the precision was ≤ 5%. The detection limit and quantification limit for Pb determination were 3.2 and 10.7 μg L − 1 , respectively. The proposed method was applied for the determination of Pb in water samples from a river heavily polluted by industrial effluents. The recovery measured by analyte addition technique showed that the proposed pre-concentration method had good accuracy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.